import importlib
import time
import numpy as np
import re
import autodeer.tools as tools
from scipy.optimize import minimize_scalar, curve_fit
from autodeer.hardware import XeprAPILink
from deerlab import correctphase
from numpy.polynomial import Polynomial
[docs]
MODULE_DIR = importlib.util.find_spec('autodeer').submodule_search_locations[0]
# =============================================================================
[docs]
def get_nutations(api, nu, field, step, ELDOR: bool = True, nx: int = 128):
min_freq = nu[0]
max_freq = nu[1]
freq_table = np.arange(min_freq, max_freq, step)
n = len(freq_table)
if len(field) == 2:
input_freq = field[0]
input_field = field[1]
start_field = input_field * min_freq / input_freq
elif len(field) == 1:
start_field = field
field_table = freq_table * start_field/min_freq
# go to start field / freq
api.set_field(field_table[0], hold=True)
api.set_freq(freq_table[0])
if ELDOR:
api.set_ELDOR_freq(freq_table[0])
nut_data = np.zeros((n, nx+1), dtype=np.complex64)
tools.progress_bar_frac(0, n)
for i in range(0, n):
api.set_field(field_table[i], hold=True)
api.set_freq(freq_table[i])
if ELDOR:
api.set_ELDOR_freq(freq_table[i])
api.run_exp()
while api.is_exp_running():
time.sleep(0.5)
dataset = api.acquire_dataset()
nut_data[i, 0] = api.get_counterfreq()
nut_data[i, 1:] = dataset.data
t = dataset.axes
tools.progress_bar_frac(i+1, n)
return t, nut_data
# =============================================================================
[docs]
def CP_run(
api, d0, num_pulses=3, ps_length=16, sweeps=4, dt=100, num_points=256,
srt = 6e6):
if num_pulses == 2:
run_general(
api,
["/PulseSpel/HUKA_DEER_AWG"],
["5p DEER relax", "DEER run AWG -+<x>"],
{"PhaseCycle": True},
{"p0": ps_length, "p1": ps_length, "h": 10, "n": sweeps, "d30": dt,
"d0": d0, "dim10": num_points, "srt": srt},
False)
else:
raise ValueError("Only CP2 is currently implemented")
# =============================================================================
[docs]
def DEER5p_run(
api, ps_length, d0, tau2, sweeps=4, deadtime=80, dt=16,
num_points=0, srt=6e6):
if num_points == 0:
num_points = np.floor((tau2 + tau2 - deadtime)/dt)
run_general(
api,
["/PulseSpel/HUKA_DEER_AWG"],
["5p DEER", "DEER run AWG -+<x>"],
{"PhaseCycle": True, "ReplaceMode": False},
{"p0": ps_length, "p1": ps_length, "h": 20, "n": sweeps, "d2": tau2,
"d11": 200, "d3": deadtime, "d30": dt, "d0": d0,
"dim8": num_points, "srt": srt},
run=False)
# =============================================================================
[docs]
class DEER:
def __init__(self, api, d0, det_frq, pump_frq, srt=6e6) -> None:
[docs]
self.pump_frq = pump_frq
pass
[docs]
def run_5p(
self, tau1: float, tau2: float, dt: float = 16,
deadtime: float = 80, num_points: int = 0,
scans: int = 200) -> None:
ps_length = 32
if num_points == 0:
num_points = np.floor((tau2 + tau2 - deadtime)/dt)
# Add some selection based on AWG / HYBRID/ MPFU
if self.awg:
print("Bruker AWG are not yet supported")
elif self.hybrid:
ps_file = "/PulseSpel/HUKA_DEER_AWG"
ps_exp = "5pDEER"
ps_pc = "DEER run"
elif self.mpfu:
ps_file = "/PulseSpel/aD_DEER_MPFU"
ps_exp = "5p DEER"
ps_pc = "DEER run AWG -+<x>"
run_general(
self.api,
[ps_file],
[ps_exp, ps_pc],
{"PhaseCycle": True, "ReplaceMode": False},
{"p0": ps_length, "p1": ps_length, "h": 20, "n": scans,
"d2": tau2, "d11": 200, "d3": deadtime, "d30": dt, "d0": self.d0,
"dim8": num_points, "srt": self.srt},
run=True)
pass
[docs]
def run_4p(
self, tau1: float, tau2: float, dt: float = 16,
deadtime: float = 80, num_points: int = 0,
scans: int = 200) -> None:
ps_length = 32
if num_points == 0:
num_points = np.floor((tau2 + tau2 - deadtime)/dt)
# Add some selection based on AWG / HYBRID/ MPFU
if self.awg:
print("Bruker AWG are not yet supported")
elif self.hybrid:
ps_file = "/PulseSpel/HUKA_DEER_AWG"
ps_exp = "4pDEER"
ps_pc = "DEER run"
elif self.mpfu:
ps_file = "/PulseSpel/aD_DEER_MPFU"
ps_exp = "4-DEER AWG"
ps_pc = "DEER run AWG -+<x>"
run_general(
self.api,
[ps_file],
[ps_exp, ps_pc],
{"PhaseCycle": True, "ReplaceMode": False},
{"p0": ps_length, "p1": ps_length, "h": 20, "n": scans,
"d2": tau2, "d11": 200, "d3": deadtime, "d30": dt, "d0": self.d0,
"dim8": num_points, "srt": self.srt},
run=True)
pass
[docs]
def run_CP(self, tau: int, dt: int = 50, num_points: int = 200,
scans: int = 1) -> None:
ps_length = 32
if self.awg:
print("Bruker AWG are not yet supported")
elif self.hybrid or self.mpfu:
ps_file = "/PulseSpel/HUKA_DEER_AWG"
ps_exp = "4pDEER"
ps_pc = "DEER run"
run_general(
self.api,
[ps_file],
[ps_exp, ps_pc],
{"PhaseCycle": True, "ReplaceMode": False},
{"p0": ps_length, "p1": ps_length, "h": 20, "n": scans,
"d1": tau, "d30": dt, "d0": self.d0,
"dim8": num_points, "srt": self.srt},
run=True)
pass
# =============================================================================
[docs]
class MPFUtune:
"""
Tuning MPFU channels for optimal attenuation and phase
"""
def __init__(self, api, echo="Hahn", ps_length=16, d0=680, srt=6e6) -> None:
"""
Parameters
----------
api : _type_
The spectrometr API object
echo : str, optional
The echo type. Options = ['Hahn","Refocused"], by default "Hahn"
ps_length : int, optional
The length of the pi/2 pulse, by default 16
d0 : int, optional
The approximate position of d0, this should be lower than ideal,
by default 680
"""
[docs]
self.hardware_wait = 5 # seconds
[docs]
self.ps_length = ps_length
if echo == "Hahn":
self._setup_echo("Hahn Echo", tau1=400)
elif echo == "Refocused":
self._setup_echo("Refocused Echo", tau1=200, tau2=400)
else:
raise ValueError(
"Only Hahn and Refocused echo's are currently supported")
pass
[docs]
def _setup_echo(self, echo, tau1=400, tau2=400):
PulseSpel_file = "/PulseSpel/phase_set"
run_general(self.api,
[PulseSpel_file],
[echo, "BrXPhase"],
{"PhaseCycle": False},
{"p0": self.ps_length*2, "p1": self.ps_length, "h": 20,
"n": 1, "d0": self.d0, "d1": tau1, "d2": tau2, "pg": 128,
"srt": self.srt},
run=False
)
[docs]
def tune_phase(
self, channel: str, target: str, tol=0.1, maxiter=30) -> float:
"""Tunes the phase of a given channel to a given target using the
standard scipy optimisation scripts.
Parameters
----------
channel : str
The chosen MPFU channel. Options: ['+<x>', '-<x>', '+<y>', '-<y>']
target : str
The target echo position, this can either be maximising (+) or
minimising (-) either the real (R) or imaginary (I) of the echo.
Options: ['R+', 'R-', 'I+', 'I-']
tol : float, optional
The tolerance in phase parameter, by default 0.1
maxiter : int, optional
The maximum number of iterations in the optimisation, by default 30
Returns
-------
float
The optimal value of the phase parameter
"""
channel_opts = ['+<x>', '-<x>', '+<y>', '-<y>']
phase_opts = ['R+', 'R-', 'I+', 'I-']
if channel not in channel_opts:
raise ValueError(f'Channel must be one of: {channel_opts}')
if target not in phase_opts:
raise ValueError(f'Phase target must be one of: {phase_opts}')
if channel == '+<x>':
phase_channel = 'BrXPhase'
elif channel == '-<x>':
phase_channel = 'BrMinXPhase'
elif channel == '+<y>':
phase_channel = 'BrYPhase'
elif channel == '-<y>':
phase_channel = 'BrMinYPhase'
if target == 'R+':
test_fun = lambda x: -1 * np.real(x)
elif target == 'R-':
test_fun = lambda x: 1 * np.real(x)
elif target == 'I+':
test_fun = lambda x: -1 * np.imag(x)
elif target == 'I-':
test_fun = lambda x: 1 * np.imag(x)
lb = 0.0
ub = 100.0
def objective(x, *args):
# x = x[0]
self.api.hidden[phase_channel].value = x # Set phase to value
time.sleep(self.hardware_wait)
self.api.run_exp()
while self.api.is_exp_running():
time.sleep(1)
data = self.api.acquire_scan()
v = data.data
val = test_fun(np.sum(v))
print(f'Phase Setting = {x:.1f} \t Echo Amplitude = {-1*val:.2f}')
return val
output = minimize_scalar(
objective, method='bounded', bounds=[lb, ub],
options={'xatol': tol, 'maxiter': maxiter})
result = output.x
print(f"Optimal Phase Setting for {phase_channel} is: {result:.1f}")
self.api.hidden[phase_channel].value = result
return result
[docs]
def tune_power(
self, channel: str, tol=0.1, maxiter=30,
bounds: list[float] = [0, 100]) -> float:
"""Tunes the attenuator of a given channel to a given target using the
standard scipy optimisation scripts.
Parameters
----------
channel : str
The chosen MPFU channel. Options: ['+<x>', '-<x>', '+<y>', '-<y>']
tol : float, optional
The tolerance in attenuator parameter, by default 0.1
maxiter : int, optional
The maximum number of iterations in the optimisation, by default 30
Returns
-------
float
The optimal value of the attenuator parameter
"""
channel_opts = ['+<x>', '-<x>', '+<y>', '-<y>']
if channel not in channel_opts:
raise ValueError(f'Channel must be one of: {channel_opts}')
if channel == '+<x>':
atten_channel = 'BrXAmp'
elif channel == '-<x>':
atten_channel = 'BrMinXAmp'
elif channel == '+<y>':
atten_channel = 'BrYAmp'
elif channel == '-<y>':
atten_channel = 'BrMinYAmp'
lb = bounds[0]
ub = bounds[1]
def objective(x, *args):
self.api.hidden[atten_channel].value = x # Set phase to value
time.sleep(self.hardware_wait)
self.api.run_exp()
while self.api.is_exp_running():
time.sleep(1)
data = self.api.acquire_scan()
v = data.data
val = -1 * np.sum(np.abs(v))
print(f'Power Setting = {x:.1f} \t Echo Amplitude = {-1*val:.2f}')
return val
output = minimize_scalar(
objective, method='bounded', bounds=[lb, ub],
options={'xatol': tol, 'maxiter': maxiter})
result = output.x
print(f"Optimal Power Setting for {atten_channel} is: {result:.1f}")
self.api.hidden[atten_channel].value = result
return result
[docs]
def tune(self, channels: dict, tol: float = 0.1,
bounds=[0, 100]) -> None:
"""Tunes both the power and attenuation for a collection of channels.
Parameters
----------
channels : dict
A dictionary of MPFU channels to be tunned and the associated phase
target.\\
Channel options = ['+<x>', '-<x>', '+<y>', '-<y>']\\
Phase target options = ['R+', 'R-', 'I+', 'I-']\\
E.g. {'+<x>': 'R+','-<x>': 'R-'}
tol : float, optional
The tolerance for all optimisations, by default 0.1
"""
for channel in channels:
if channel == '+<x>':
phase_cycle = 'BrXPhase'
elif channel == '-<x>':
phase_cycle = 'BrMinXPhase'
elif channel == '+<y>':
phase_cycle = 'BrYPhase'
elif channel == '-<y>':
phase_cycle = 'BrMinYPhase'
self.api.set_PulseSpel_phase_cycling(phase_cycle)
print(f"Tuning channel: {channel}")
self.tune_power(channel, tol=tol, bounds=bounds)
self.tune_phase(channel, channels[channel], tol=tol)
[docs]
def calc_d0(self):
initial_d0 = 500
PulseSpel_file = "/PulseSpel/phase_set"
run_general(
self.api,
[PulseSpel_file],
['Hahn Echo Trans', "BrXPhase"],
{"PhaseCycle": False},
{"p0": self.ps_length*2, "p1": self.ps_length, "h": 20, "n": 1,
"d1": 400, "d0": initial_d0}
)
while self.api.is_exp_running():
time.sleep(1)
data = self.api.acquire_scan()
max_pos = np.argmax(np.abs(data.data))
max_time = data.axes[max_pos]
d0 = initial_d0 + max_time
return d0
# =============================================================================
[docs]
class ELDORtune:
def __init__(self, api: XeprAPILink, d0=700, ps_length=16, srt=6e6) -> None:
"""
Tuning incoherent ELDOR channel for optimal power using nutation
experiments
Parameters
----------
api : XeprAPILink
The spectrometer API object
d0 : int, optional
The approximate position of d0, this should be lower than ideal,
by default 700
ps_length : int, optional
The length of the pi/2 pulse, by default 16
"""
[docs]
self.ps_length = ps_length
pass
[docs]
def _setup_exp(self, tau1=400, tau2=400):
PulseSpel_file = "/PulseSpel/param_opt"
run_general(self.api,
[PulseSpel_file],
["nutation", "ELDOR nut"],
{"PhaseCycle": True},
{"p0": self.ps_length*2, "p1": self.ps_length, "h": 20,
"n": 1, "d0": self.d0, "d1": tau1, "d2": tau2, "pg": 128,
"dim7": 32, "srt": self.srt},
run=False
)
[docs]
def _get_exp(self):
self.api.run_exp()
data = self.acquire_scan()
return data
[docs]
def find_min(self, dataset):
data = correctphase(dataset.data)
data /= data.max()
time = dataset.axes
def test_fun(x, A, B, freq, phase, decay):
return A*np.sin(x*2*np.pi*freq + phase)*np.exp(-1*x*decay) + B
bounds = (
[-1.5, -1, 0, -np.pi, 0],
[1.5, 1.5, 1, np.pi, 10]
)
p0 = [0.5, 0.6, 0.04, 1.4, 0.03]
popt, pcov = curve_fit(
test_fun, time, data, p0=p0, bounds=bounds)
return 1/(2*popt[2])
[docs]
def tune(self, target: int):
lb = 0
ub = 30
tol = 2
maxiter = 10
def objective(x, *args):
self.api.set_attenuator("ELDOR", x)
time.sleep(self.hardware_wait)
self.api.run_exp()
while self.api.is_exp_running():
time.sleep(1)
dataset = self.api.acquire_scan()
min = self.find_min(dataset)
print(f'Atten Setting = {x:.1f} \t pi pulse time = {min:.1f} ns')
return abs(min-args[0])
output = minimize_scalar(
objective, method='bounded', bounds=[lb, ub],
options={'xatol': tol, 'maxiter': maxiter}, args=target)
result = output.x
print(f"Optimal ELDOR atten is: {result:.1f}")
self.api.set_attenuator("ELDOR", result)
return result
# =============================================================================
[docs]
class PulseProfile:
def __init__(self, api: XeprAPILink, d0=700, ps_length=16, srt=4e6) -> None:
"""
Tuning incoherent ELDOR channel for optimal power using nutation
experiments
Parameters
----------
api : XeprAPILink
The spectrometr API object
d0 : int, optional
The approximate position of d0, this should be lower than ideal,
by default 700
ps_length : int, optional
The length of the pi/2 pulse, by default 16
"""
[docs]
self.ps_length = ps_length
pass
[docs]
def _setup_exp(self, tau=400):
r"""
Setup the pulse profile experiment.
Parameters
----------
tau : int, optional
The seperation between :math:'\pi/2' and :math:'pi' in the Hahn
echo, by default 400
"""
PulseSpel_file = "/PulseSpel/param_opt"
run_general(self.api,
[PulseSpel_file],
["PulseProfile ELDOR", "ELDOR nut"],
{"PhaseCycle": True},
{"p0": self.ps_length*2, "p1": self.ps_length, "h": 20,
"n": 1, "d0": self.d0, "d1": tau, "pg": 128,
"dim8": 10, "srt": self.srt},
run=False
)
[docs]
def _freq_sweep(self, nu: list, step: float, gyro: float):
"""
Run the frequency sweep for a pulse profile.
Parameters
----------
nu : list
A list detailing the starting and ending frequency,
[nu_init, nu_final]
step : float
The frequency step, given in GHz
gyro : float
The gyromagnetic ratio in G/GHz.
Returns
-------
_type_
_description_
"""
min_freq = nu[0]
max_freq = nu[1]
freq_table = np.arange(min_freq, max_freq, step)
n = len(freq_table)
PP_x: np.ndarray = np.zeros(n, dtype=np.float32)
PP: np.ndarray = np.zeros(n, dtype=np.complex128)
# go to start field / freq
self.api.set_freq(freq_table[0])
self.api.set_field(self.api.get_counterfreq()/gyro, hold=True)
tools.progress_bar_frac(0, n)
for i in range(0, n):
self.api.set_freq(freq_table[i])
self.api.set_field(self.api.get_counterfreq()/gyro, hold=True)
self.api.run_exp()
while self.api.is_exp_running():
time.sleep(0.5)
dataset = self.api.acquire_dataset()
PP_x[i] = self.api.get_counterfreq()
PP[i] = np.mean(dataset.data)
tools.progress_bar_frac(i+1, n)
return PP_x, PP
# =============================================================================
[docs]
def CalibrateFreq(api: XeprAPILink, num_points: int = 50, deg: int = 5):
"""Generate the polynomial parameters for converting from frequency
(in GHz) to Xepr gunn diode stepper value. 0-4095.
Parameters
----------
api : XeprAPILink
The API for the spectrometer
num_points : int, optional
The number of points to be measured, by default 50
deg : int, optional
The degree of polynomial fit, by default 5
"""
positions = np.linspace(0, 4095, num_points, endpoint=True)
array = np.zeros((2, positions))
for i, val in enumerate(positions):
array[0, i] = val
api.hidden['Frequency'] = val
time.sleep(0.5)
array[1, i] = api.get_counterfreq()
pfit = Polynomial.fit(array[1, :], array[0, :], deg)